(1)旋下压力容器上的注水嘴的螺栓,在工作台面上同时打开7只阀门,将洁净的清水注入压力容器内,观察0号控制阀放水管出水流畅,然后关闭0号控制阀,直至注满溢出为止,再将螺栓装上拧紧。
(2)向进水箱注满清水。
(3)将水泵上二异柱螺母拆下,拔出上挡板及水泵柱塞,向泵内注满清水,再按原方向装上即能开机正常工作。
(4)调节电接点压力表至1.5兆帕位置上。
(5)通电,启动电源开关,使水泵能正常运行(应无冲击现象)观察6个试模座出水是否正常,直至无气泡为止。
(6)逐个关闭控制阀,观察压力表压力是否上升,排除一切泄漏故障,使压力能稳定在所调节的范围内。
(7)按标准试验方法,将试模逐个安装就位,初步不可拧紧,打开各个试模控制阀门,观察各个试模座泄出水为止,拧紧螺母,调整好欲按要求试验压力的电接点压力表,即可正常工作。